Polaris FID – THC Analysis

 

The POLARIS FID from Italian Company Pollution SRL is an on-site portable TOC Analyser for Stack Emission Monitoring in compliance with EN12619 and EN13526 and USEPA method 25A.  It is a fully portable heated FID instrument and it incorporates everything needed for full analysis onsite.

 

The miniaturization of the flame ionization detector (FID) and volumetric sampling system result in outstanding energy savings and thus allows the use of built-in rechargeable batteries. It also ensures the highest accuracy and repeatability of sequence analysis.

The compact hydrogen storage cartridge (using an innovative metal-hydride technology) is integrated into the instrument and self-desorbs the hydrogen fuel to the flame detector at appropriate pressure and flow. It is very safe and it does not require pressure regulators; it is a very handy instrument because it allows to perform continuous analysis for many hours.

Because of its peculiarities, it can be carried by car and plane without any restrictions. It can be easily recharged in a few hours in the laboratory using a standard hydrogen tank with regulator or with a hydrogen generator at high output pressure.

The instrument is designed to be resistant to samples with high temperature and humidity: this is a typical working condition with hot samples during stack monitoring. The detector manifold, the sampling lines and all the valves are uniformely heated and integrated into a single block, in order to eliminate all possible cold spots and avoiding local sample condensation.

The instrument incorporates everything you need for analysis: batteries, hydrogen storage accumulator, pump and Activated Charcoal filter for the flame air, span gas cylinder for the calibration and technical air cylinder for high accuracy and precision. The sample inlet, with double filter, is compatible with all sampling lines on the market. It is equipped with comfortable shoulder strap for field use, also available as a front-bag.

The integrated microprocessor automatically controls several parameters: ignition and flame monitoring, setup of parameters chose by the user according to the analytical method, diagnostic checks, life monitoring of the gas cans and energy reserves, sampling lines cleaning cycles, calibration procedures.

 

Key Features

Technical Data

  • Size And Weight : 355(W) x 155(H) x 426(D) mm, 10 Kg (with batteries 13,4 Kg)
  • Power Supply : External Power Adapter 240VAC – 24VDC @ 8,2A
  • Sampling : Heated head system
  • Environmental Operating Ranges : +5°C < T < +40°C – 0% < RH < 95%
  • Sample Conditions : Temperature 0 – 170°C / pressure 90 – 110kPa
  • Max Detector Body Temperature : 190°C
  • Max Temperature Sample Inlet : 180°C
  • Materials In Contact With The Sample : Steel AISI316 and high performance polymers
  • Sample Flow : approximately 800 ml/min
  • Sample Pressure : atm ± 100 mbar
  • Measuring Range : 0-20/50/150/500/5000 mg/Nm3
  • Detection Limit (LDL) : 0,3 mg/Nm 3
  • Accuracy : 2% of F.S.
  • Repeatability : 2% of F.S.
  • Setting Methods And Data Analysis : Internal memory flash (30 days of continuous measurements) – Upload/download via USB
  • Flame Management : Electronic, with diagnostic and ‘flame off’ visual alarm on screen, automatic system to restart the flame
  • Hydrogen Storage : Cartridge, rechargeable via external source of hydrogen under moderate pressure. No restriction transport by car or plane
  • Zero Air : 1L Cylinder (non-rechargeable): technical air @ 12bar / 20°C (running time approximately 10h). Connectors for external Zero Air cylinder
  • Span Gas : 1L Cylinder (non-rechargeable): propane in Air @ many concentrations – Quick connectors for external Span Gas cylinder

 

 

 

 